Aerosol Diluter ATI – TDA D10 / TDA D100
. Compatible with any 1 cfm particle counter.
. Minimizes pressure drops and aerosol loss.
. Available in 10:1 and 100:1 dilution ratio configurations.
. Can be plumbed in series for alternate dilution configurations.
. Ideal for isolators, mask fittings and installation with low airflows.
. Works with any type of aerosol challenge, including PSL, PAO, DOP, and NaCl.
Product Specifications
Dilutes aerosol to measurable levels for particle counters.
Installed between the aerosol source and a particle counter, diluters minimize pressure drop and aerosol loss while preventing saturation of the particle counter.
Dilution Ratio: D10 – 10:1 | D100 – 100:1
Particle Size of Interest: 0.1, 0.3 or 0.5 micron
Air Flow Rate: 1 cfm (28.3 lpm)
Pressure Drop: Nominally ≤ 1 in. Water Column (25 mm WC)
Electrical: Not required
Run Time Meter: Included
Material: Stainless Steel
Operational Inlet / Outlet: 0.325 in. Tubing ID | 0.25 in. Tubing ID Optional
Size: 15.5 cm x 8.3 cm x 55.9 cm (6.5 in. x 3.3 in. x 22.0 in.)
Weight: 3.0 kg (7.0 lb)
Maintenance: Replace internal bypass filters as necessary. Ultrasonic cleaning of annular orifice. Clean or replace tubing as required.
* Note: Flow rate and particle size of interest can be configured to customer requirements
Application:
Pharmaceutical manufacturers.
Medical facilities and cleanrooms.
Associated services
. Training in the use of the device, remotely or on site, and in the use of the data.
. Performance of metrological verification services in our laboratory located in Betton (35).
You may also like…
-
Photometer ATI Model TDA 2i
. Hepa Filter Testing ISO 14644-3.
. Lightweight, portable and rugged digital photometer
Large.
. 4.3″ LCD screen to display critical information.
. iProbe (optional) with 12 ft cable for hard to reach areas minimizes downtime.
. User settable aerosol noise suppression for more stable aerosol measurements.
. Printer (optional) available for reporting. -
Thermal Aerosol Generator ATI Model TDA 5D
. Hepa Filter Testing ISO 14644-3.
. Fast operation with a warm-up time of only 60 seconds.
. Widest range of System Flow Rates 850 – 120.000 m3/hr.
. Rugged Portability.
. Safe and reliable Operation.
. Flex-Fuel Capacity.
. Lightweight and rugged, ideal for remote locations.