Microbial Air Sampler CLIMET Model CI-90A
Features
Configurations:
• CI-90A = 100 LPM microbial sampler with Aluminum Sampling Head.
• CI90A+ = 100 LPM microbial sampler with Passivated 316L Stainless Steel Sampling Head.
* All models have 304 Stainless Steel Enclosure, RS-232 data interface, and HEPA filtered internal exhaust. Both sampling heads are autoclavable @ 134oC for 20 minutes.
Performance:
– Sample Flow Rate: 100 LPM (0.1 cubic meter per minute, or 3.53 cubic feet per minute).
– Sample time to measure a cubic meter: 10 minutes.
– Flow Rate Control: Electronic, automatic closed loop.
– Cleaning: Compatible with common cleaning and sterilization procedures, including VHP.
– Flow Rate Audible Alarm: Internal audible alarm, +/-10% of 100 LPM Flow Rate.
– Data Interface: RS-232 for use with IDLoader Sofware.
– Display: 4×20 character LCD.
– Printer: Thermal Label Printer for traceability (Location ID, User ID, time, date, etc).
– Programming & Memory Unit ID: User Selectable; 6 characters long.
– User ID: Up to 10 users, 9 characters long.
– Location ID: Up to 100; 15 characters.
– Security: Prevents changes to settings, or stopping a sample in progress.
Included Accessories and Features:
Internal thermal label printer, internal battery charger, battery, stainless steel locking collar, sample head, two rolls of adhesive labels, manual, and IDLoader Sofware included.
Electrical AC and battery supported:
– Battery: 9.6 VDC, 7.6 Amp Hours.
– Battery Operation (typical): 7 hours with continuous sampling.
– Battery Recharge time: 2.0-2.5 hours (built-in charger).
– Wall Power: Universal 100-240 VAC Input.
Dimensions & Weight: (19.05 cm x 20.574 cm x 25.146 cm).
CI-90A with aluminum sample head, weight with battery is 4.76 kg.
CI-90A+ with stainless steel sample head, weight with battery is 5.44 kg.
Environmental:
Operating temperature: 0-36⁰C, 32-97⁰F.
0-90% relative humidity, non-condensing.
• Biological and Physical Efficiency validated to ISO 14698-1, with unparalleled performance down to 1μm. Biological and Physical Efficiency test report available to customers upon request.
Accessories
. Stainless steel sample head
. Remote sampling and compressed gas accessories
. Carrying case
Associated services
. Equipment rental: available within 48 hours (depending availability). Delivered with all necessary accessories, calibration certificate and user manuel.
. IQ/OQ customised protocols to your application.
. On site training and support during installation and set up of program parameters and SOP.
. Calibration and repair in our laboratory located at Rennes (35).